1: Generator
Output power: 5kW
Inverter frequency: 40kHz
Continuous fluoroscopy
Pulse fluoroscopy
Radiography:
-Tube current: 25mA ~ 100mA
2: Tube: double focus -0.6/1.8
3: Image intensifier: Canon 9*9 inch
4: Built-in C-arm frame (24” monitor)
Power supply: single phase 220V 50/60Hz
Optional DAP
Compact design digital workstation Special professional software for veterinary use, intelligent and compact design with built-in computer workstation, space saving and convenient to use.
Hand controller Humanized miniature, movable operator panel design, free control from host, practical and convenient.
Image intensifier Toshiba brand image intensifier with reliable quality along with dense grid, improving superior image acquisition
X-ray tube
X-ray tube combination assembly with high output power, cable-free design, beautiful structure, easy to maintain
Doctors’ eyes
Smart design with flexible movement, provides a larger operating space.
Fluoroscopic Capacity
Max rated capacity Tube Current 4mA, Tube Voltage 120kV
Automatic Fluoroscopy Tube Voltage:40kV~120kV
Manual Fluoroscopy Tube Current : 0.3mA~4mA
Continuous Tube Voltage:40kV~120kV
Continuous Tube Current : 0.3mA~4mA
Continuous Tube Voltage 40kV~120kV
Continuous Tube Current 4.1mA~30mA
Pulse frequency 0.1~25fps
Photography Capacity
Max rated capacity 5 KW
Tube Voltage and Current Combination 40kV~49kV 1 ~180 mAs
50kV~59kV 1 ~140 mAs
60kV~69kV 1 ~125 mAs
70kV~79kV 1 ~110 mAs
80kV~89kV 1 ~100 mAs
90kV~99kV 1 ~80 mAs
100kV~109kV 1 ~63mAs
110kV~120kV 1 ~50mAs